Do note that this checklist, although fairly comprehensive, does not cover the subtleties of the different types of loggers. We recommend that you contact our team, with a combined experience of over 70 years in the temperature measuring field, to get advice and recommendations, specific to your needs.

This table is quite large, and thus has been split into two.

Table 1: Comparing the T-TEC E, T-TEC F, T-TEC G and T-TEC P data loggers

ChecklistT-TEC ET-TEC FTempTec 816
 
T-TEC P
Temperature Range

T1 - 40 + 85 C
T2 -100 + 20 C 
T6 +20 + 160 C (T-TEC 6 model only)

T1 - 40 + 85 C 
T2 -100 + 20 C 
T6 +20 + 160 C (T-TEC 6 model only)
- 40 to +90 CA -200 + 120 C 
B -40 + 120 C 
H -40 + 300 C
Accuracy

T1

-40 to 0 C 0.3 C 
0 to 70 C 0.2 C 
70 to 85 C 0.3 C 
85 to 100 C 0.5 C

T2

-90 to 20 C 0.8 C 
-100 to -90 C 1.0 C

Time: Approx. 1 minute per month

T1

-40 to 0 C 0.3 C 
0 to 70 C 0.2 C 
70 to 85 C 0.3 C 
85 to 100 C 0.5 C

T2

-90 to 20 C 0.8 C 
-100 to -90 C 1.0 C

Time: Approx. 1 minute per month

-40 to 0 C 0.3 C 
0 to 70 C 0.2 C 
70 to 85 C 0.3 C

Time: Approx. 1 minute per month

Please see Sensor Specifications page as each different sensor has different accuracy over their range.

Time: Approx. 1 minute per month

Resolution< 0.1 C< 0.1 C< 0.1 C< 0.1 C
Water Resistancy

IP68, Dust and water tight, IEC standard 60529

IP68, Dust and water tight, IEC standard 60529No ratingIP68, Dust and water tight, IEC standard 60529
Length of RunDetermined by sample rateDetermined by sample rateDetermined by sample rateDetermined by sample rate
Sample Rate

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Memory12, 000+ logs10, 600 logs per channel 
21, 200 logs if only 1 sensor used
20, 600 logs20, 000 to 60,000 logs
Display Options

Type 7 has LCD

Type 6 has none

Type 7 has LCD

Type 6 has none

LCD

Type 7 has LCD

Type 6 has none

Output optionsOptional Relay, Audible alarms (Type 6), Visual Alarms (Type 7)Audible allarms (Type 6), Visual Alarms (Type 7)Two possible outputs, visual alarms, audible alarms, LCDAudible allarms (Type 6), Visual Alarms (Type 7)
Physical Size65 31 106 mm65 31 106 mm

118mm 88mm 60mm

(L D H)

65 31 106 mm
Number of inputs1211
Starting method

Immediate

Delay

Immediate

Delay

Immediate

Delay

Immediate

Delay

Reading MethodConnection to computer via USB or serial portConnection to computer via USB or serial portConnection to computer via USB or serial portConnection to computer via USB or serial port
ConstructionPolycarbonate (food grade, impact resistant)Polycarbonate (food grade, impact resistant)Polycarbonate (food grade, impact resistant)/ABSPolycarbonate (food grade, impact resistant)

 

Table 2: Comparing the T-TEC A, T-TEC C and thermochron data loggers

ChecklistT-TEC AT-TEC CiButton
Temperature Range4-20mA input

- 40 + 85 C

0-100% RH

DS1921G F52
- 40 + 85 C

DS1921H F5
+ 15 + 46 C

DS1921Z F5
- 5 + 25 C

DS1922E
+ 15 C to +140 C

DS1922L
- 40 C to +85 C

DS1922T
0 C to +125 C

DS1923
- 20 C to +85 C

Accuracy

Better than 1% of full scale

Time: Approx. 1 minute per month

-40 to 0 C 0.3 C 
0 to 70 C 0.2 C 
70 to 85 C 0.3 C 
85 to 100 C 0.5 C

0-100 % 2% RH, Compensated for -5 + 60 C ambient temperature

Time: Approx. 1 minute per month

1 C from -20 + 70 C, at time of calibration at factory.

Time: Maximum 8 minutes per month over full temperature range.

DS1922E: 1.5 C from +110 C to +140, 7 C typical from +15 C to +110 C

DS1923: 5% RH

Resolution10 A

Better than 0.1 C

Better than 0.1%RH

DS1921L F52: 0.5 C 
DS1921H F5: 0.125 C 
DS1921Z F5: 0.125 C

DS1922E/L/T: 0.5 C or 0.0625 C (low/high res.)

 

DS1923: 0.5 C, 0.64%RH or 0.0625 C , 0.04%RH (low/high res.)

Water ResistancyIP68, Dust and water tight, IEC standard 60529Splash proof from top. In order to measure humidity, water vapor does need to reach the sensor.IP65 (Water resistant only)
Length of RunDetermined by sample rateDetermined by sample rateDetermined by sample rate
Sample Rate

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Seconds: 1, 2, 5, 10, 30

Minutes: 1, 2, 5, 10, 30

Hours: 1, 2, 6

Minutes: 1, 2, 5, 10, 15, 30

Hours: 1, 2

Memory20, 000 to 60,000 logs10,600 sets each sensor

DS1921: 2, 048 logs
DS1922: 4,096 or 8,192 logs (dep. resolution)
DS1923: 2,048 or 4, 096 (dep. resolution)

Display Options

Type 7 has LCD

Type 6 has none

Type 7 has LCD

Type 6 has none

None
Output Options

Type 6: audible alarms

Type 7: visual alarms, LCD

Type 6: audible alarms

Type 7: visual alarms, LCD

None
Physical Size65 31 106 mm65 31 106 mm

17.35mm 5.9

(Diameter Height)

Number of inputs11 Temperature, 1 humidity1
Starting Method

Immediate

Delay

Immediate

Delay

Immediate

Delay

Reading MethodConnection to computer via USB or serial portConnection to computer via USB or serial portConnection to computer via USB port
ConstructionPolycarbonate (food grade, impact resistant)Polycarbonate (food grade, impact resistant)Stainless Steel, Polypropylene seal

Notes

Construction

Polycarbonate is stable in water, alcohol, oil, salts and weak acids but adequate care should be taken as it can become cloudy, swell or dissolve in alkali, armoatic hydrocarbons, halogenated hydrocarbons, etc. Printing on the casing may be damaged by alcohol.

Polycarbonate is long time temperature stable up to 115 C. It is highly weather resistant and may be left outdoors for prolonged time with little effect on properties.

The sanitary properties are good and meet the sanitary standards of FDA (USA) and is acceptaed for contact with food by the Japanese health authorities.

The casing's resistance to some chemicals is reduced at elevated temperatures (>60 C) . Please enquire if in doubt.

Cleaning: Either soap and water. Alcohol can be used with care but may damaged printing on the casing. IP68 loggers can be placed into a dishwasher.

Accuracy

The accuracy listed is for both the sensor and unit combined, not the unit singly.

Send mail to Webmaster with questions or comments about this web site.